gel electrophoresis
A. gel casting tray manufactured in UV transparent acrylic. End grooves for seating the gel casting gates. Slots for positioning the combs
B. On the Maxi and Maxi Plus models, 5 mm gates fit neatly into the grooves to create convenient tape free seals.
C. Pure platinum electrodes are easily replaced by the user.
D. Carrying handles add strength to the units and facilitate moving. - to empty the buffer for example.
E. When the lid is open a physical barrier prevents the power leads from making electrical contact with the tank.
F. A white gel table enhances the visibility of market dyes and samples as they migrate through the gel.
G. A wide range of combs, color coded for thickness.
H. When attached to the white gel table, colored strips enhance visibility when sample loading.
Gel Casting for Mini and Midi series.
Removable UV transparent gel casting tray with end gaskets which allow the gel to be cast directly in the running tank
Lift the gel running tray out of the electrophoresis unit. Make sure that the silicone gasket is seated evenly in the groove, with the ends flush with the top of the tray. Turn the tray through 90 degrees
 
Insert the tray so that the silicone rubber forms a tight, leak free seal with the walls of the gel tank
 
The agarose gel should be poured into the tray and allowed to set. When ready, lift the running tray, turn it through 90 degrees and replace in the tank
 
Gel is ready for sample loading and run.
electrophoresis fast mini
 
Gel Casting for Maxi Series
Remove the casting gates from the gel casting tray. Make sure that the silicon gaskets in the casting gates are seated evenly in the grooves, with the ends of the gaskets flush with the tops of the gates. Place the gaskets into the grooves of the gel casting tray with the silicone rubber gasket facing the outside of the tray
Insert the gates into the ends of the UV tray and the comb(s) into the comb slots
 
Agarose gel solutions should be cooled to 50 degs C, poured into the tray and allowed to set
 
When ready the casting gates and comb(s) are removed and the gel casting tray placed in the electrophoresis unit