Model |
ATR-B |
ATR-F |
Measuring scales: |
Refractive Index (RI)
Sucrose (% Brix)
10 standard scales included (Inverted Sugar (%), Fructose (%), Glucose (%), Honey (% water), Phenol (%), Acetic Acid (%), Oechsle (°Oe), Propylene Glycol (%), Zeiss, H2O2(%),
Up to 1000 scales freely definable |
Measuring range: |
1.29000 - 1.58000 I / 0 - 100 % Brix |
1.32000 - 1.7000 RI / 0 - 100 % Brix |
Resolution: |
0.00002 RI / 0.02 % Brix |
0.00001 RI / 0.01 % Brix |
Precision: |
± 0.00004 RI /
± 0.04 % Brix |
± 0.00002 RI / ±0.02 % Brix |
Reproducibility: |
± 0.00002 RI /
± 0.02 % Brix |
± 0.00001 RI / ± 0.01 % Brix |
Ambient temperature: |
+10 to +40°C |
Automatic temperature compensation: |
+5 to +80°C |
+5 to +100°C |
Temperature measurement: |
NTC sensor for measurement of sample temperature placed inside the prism |
Temperature regulation sample: |
By build-in solid state Petier-thermostat |
By build-in solid state Petier-thermostat |
Measurement mode: |
Single sample or flow through measurement |
Prism: |
YAG |
Light source / wavelength: |
LED, interference filter 589 nm |
Display: |
7" touch screen, 800 x 480 Pixel, 16 Bit colours |
Operation: |
Touchscreen, keyboard**, mouse**, barcode-reader**, remote via PC** |
Interfaces: |
1 xRS232 C serial, 3 x USB (A), 1 x USB (B), 1 x Ethernet, Easy connection of key- board, mouse, printer, bar code reader, PC and network. |
Standard model: |
ATR-B 132 |
ATR-F 342 |
Conformity: |
International Pharmacopoea, ASTM, AOAC, DIN, FDA, ICUMSA and others |