Modern - Digital -
Ergonomic Rugged reliable and easy to operate the AF200
pitot tube anemometer measures air flow in the range 1.5 to 58 fps
and pressure in the range 0 - 20mBar
Features of the AF200 Air flow meter
Key
Features • Velocity, temperature and pressure measurement • Multi-point averaging • Smoothing • Correction mode for temperature, pressure and K
factor • Easy zero
adjust • Measurements
in both imperial and metric units • Last configuration memorized • Simplicity and reliability • Choice of pitots • Optional protective cover and sling • Over 300 hours battery life • Backlit display
Kit
includes
12" pitot tube and temperature probe (top).
Optional
telescoping pitot tube - 8" to 38"
AIRFLOW PRECISION AIR VELOCITY INSTRUMENTS
The AF200
offers high standards of performance and value for precision air velocity
measurements. The pitot tube method of measurement employed by the AF200
is suitable over the range 1.5m/s to 58m/s and up to +600°C. The instrument
can be corrected for temperature, static pressure, and K factor by keying
in the appropriate values. When used with the Pt100
resistance thermometer the temperature correction can is applied automatically
up to 200C. The Averaging function produces a running average from a series
of readings as they are taken. (For example at different points accross
a duct)
A full
kit is available to accompany the AF200. This includes 295mm/12” pitot
tubes, air temperature probe, instrument protective cover and sling, quick
fit pressure connectors, rubber tubing, nylon tubing and
9V PP3 battery. All this fits neatly into a custom carrying
case.
The
instrument also functions as a stand alone micro-manometer or temperature
meter.
AF200
specification for pitot tube air velocity
Measurement Ranges
Unit
Range
Unit
Range
m/s
1.5 to 58.0
mbar
0.00 to 19.99
ft/min
300 to 9999
inH2O
0.000 to 8.029
°C
-60 to +590
mmH2O
0.0 to 203.9
°F
-76 to +1094
Pa
0 to 2000
hPA
0.00 to 20.00
Correction Ranges selected from the keypad
Value
Display Symbol
Unit
Settable in Range
Air Temperature
CORR °C
Celcius
-100 to +600
Static Pressure
CORR mbar
mbar (Absolute)
500 to 1500
Pilot Tube Cal Factor
CORR k
(factor)
0.10 to 3.00
Instrument Accuracies
Measurement
Ambient Temperature
Specification
Pressure
+15°C to +25°C / +59°F to +77°F
-10°C to +50°C / +14°F to +122°F
±0.2% Reading ±1 digit
±
0.3% Reading ±1 digit
Temperature
-10°C to +50°C / +14°F to +122°F
±0.1% Reading ±1 1°C / ±2°F
Velocity
+15°C to +25°C / +59°F to +77°F
+15°C to +25°C / +59°F to +77°F
±0.2m/s at 10m/s to 58m/s, ±40ft/min, 2000 to 9999ft/min
±0.3m/s at 4m/s, ±60 ft/min at 800 ft/min
Specification
Operating Temperature
-10°C to +50°C / +14°F to +122°F
Battery Type
9V PP3S or equivalent
Battery Life
300 hours typical (using Alkaline Battery)
Display
12.7mm / 0.5” custom LCD with automatic legend
display
Low Battery Warning
‘BAT’ shows on display
Auto Power Off
5 minutes approx., disabled during correction
and averaging modes