Thermocouple Transducer TZ-5CA

     Features
  • DC power supply 24VDC
  • DIN rain mounting
  • Input/Output/Power supply isolated
  • This terminal block-integrated temperature transducer is used together with a JIS compliant thermocouple.
  • download .pdf file

TZ-5CA Common Specifications
Base Accuracy ±0.2% F.S (25±2ºC) ±0.2% F.S
Power Supply Variation ±0.06 % F.S
Load resistance variation ±0.06% F.S
Temperature characterisitc ±0.02 % F.S/ºC
Accuracy of cold junction compensation ±1ºC (10 to 30ºC)
Disconnection detection Upside (135± 15% F.S)
Vibration resistance Frequency: 10 to 55Hz; amplitude (half): 0.15mm to 10 sweeps of 5min each in X,Y, and Z directions
Response Time 500 msec (TYP) (0-90%)
Error caused by input range setting change: within ±2 %F.S
Front adjustments ±5% for zero and span
Insulation resistance Between input and output/power supply: more than 100Mohm at 500VDC
Dielectric strength Between input and output/power supply: for 1 min at 1500VAC
Power supply voltage 24VDC ±10%
Consuming current Less than 35mA (at current output 24 VDC) //Less than 30 mA (at voltage output 24VDC)
Operating ambient temp -5 - 50ºC
Operating ambient humidity Less than 90% RH (No-condensing)
Storage Temp -10 to 70ºC
Storage Humidity Less than 60% RH (no condensing)
weight approx 80g
 
 
INPUT MODULES -
 
Code # Thermocouple Input Signal For Code Y Input Resistance Input Range
J0
J
0 to 1000ºC -50 to 1000ºC span
200 to 1050ºC
More than 1M ohm
±150%
JY  
K2
K
-50 to 200ºC -50 to 1200ºC span
200 to 1250ºC
K3 0 to 200ºC
K4 -50 to 1200ºC
K5 0 to 1200ºC
KY  
R0
R
0 to 1700ºC 0 to 1700ºC span
400 to 1700ºC
RY  
S0
S
0 to 1700ºC 0 to 1700ºC span
500 to 1700ºC
SY  
T0
T
-50 to 350ºC -50 to 350ºC span
200 to 400ºC
T1 0 to 350ºC
TY  
 
OUTPUT MODULES -
 
Code # Output Signal Allowable Load Resistance
0 0 - 5V DC More than 2 Kohm
1 1 - 5V DC
2 0 - 10V DC More than 4Kohm, negative output: more than 10Kohm
3 -10 - 10V DC
4 -2 -2VCD More than 2Kohm, negative output: more than 10Kohm
5 -2.5 - 2.5 VDC
6 -5 - 5 VDC
7 0 - 4VDC More than 2Kohm
A 4 - 20mA DC Less than 550ohm
B 0 - 20mA DC
Y Voltage ouput: less than +15 VDC and more than -12 VDC
Minimum span: less than +27 VDC and more than 0.06 VDC
Road resistance: 10Kohm at output exceeding 10V, and a negative output
Base accuracy: ±0.15 %F.S and temperature characterisitc ±0.03 %F.S/ºC for a span of less than 1V
Current output: Less than +20 mADC and more than 0 mADC
Minimum Span: Less than +20 mADC and more than 1 mADC -- outputs can be reversed for both voltage and current outputs

   

Ordering Information
TZ-5CA Configure Input/Output in Store
$288.00
 

Topac Inc. 231 CJC Highway, Suite 103 Cohasset MA 02025 USA
Tel: 781 740 8778 Fax: 781 740 8779
e-mail: sales@topac.com