Thermocouple temperature transducer(TZ-1CA)

     Features
  • AC power supply 100 - 240V AC
  • DIN rail mounting
  • Input/Output/Power supply isolated
  • download .pdf file

TZ-1CA Common Specifications
Base Accuracy
±0.2% F.S (25±2ºC)
Power Supply Variation ±0.06 % F.S
Load resistance variation ±0.06% F.S
Temperature characterisitc ±0.02 % F.S/ºC
Response Time 500 msec (0-90%)
Accuracy of cold junction compensation ±1ºC (10-30ºC)
Front Adjustments ±5% for zero and span
Insulation resistance Between input and output/power supply: more than 100Mohm at 500VDC
Dielectric strength Between input and output/power supply: for 1 min at 2000VAC
Power supply voltage 100 to 240VAC
Consuming current Less than 35mA at current output/ Less than 30mA at voltage output
Vibration resistance Frequency: 10-55Hz ; amplitude (half): 0.15mm to 10 sweeps of 5min each in X,Y, and Z directions
Operating ambient temp -5 - 50ºC
Operating ambient humidity Less than 90% RH (No-condensing)
Storage Temp -10 to 70ºC
Storage Humidity Less than 60% RH (no condensing)
weight approx 80g
 
 
INPUT MODULES -
 
Code # Thermocouple Input Signal For Code Y Input Resistance Input Allowable Range
J0
J
0 to 1000ºC -50 to 1000ºC span
200 to 1050ºC
More than 1Mohm ±150%
JY other than above -->
K2
K
-50 to 200ºC -50 to 1200ºC span
200 to 1250ºC
K3 0 to 200ºC
K4 -50 to 1200ºC
K5 0 to 1200ºC
KY -->
R0
R
0 to 1700ºC 0 to 1700ºC span
400 to 1700ºC
RY -->
S0
S
0 to 1700ºC 0 to 1700ºC span
500 to 1700ºC
SY -->
T0
T
-50 to 350ºC -50 to 350ºC span
200 to 400ºC
T1 0 to 350ºC
TY -->
 
OUTPUT MODULES -
 
Code # Output Signal Allowable Load Resistance
0 0 - 5V DC More than 2 Kohm
1 1 - 5V DC
2 0 - 10V DC More than 4Kohm, negative output: more than 10Kohm
3 -10 - 10V DC
4 -2 -2VDC More than 2Kohm, negative output: more than 10Kohm
5 -2.5 - 2.5 VDC
6 -5 - 5 VDC
7 0 - 4VDC More than 2Kohm
A 4 - 20mA DC Less than 550ohm
B 0 - 20mA DC
Y Voltage ouput: less than +15 VDC and more than -12 VDC
Minimum span: less than +27 VDC and more than 0.06 VDC
Road resistance: 10Kohm at output exceeding 10V, and a negative output
Base accuracy: ±0.15 %F.S and temperature characterisitc ±0.03 %F.S/ºC for a span of less than 1V
Current output: Less than +20 mADC and more than 0 mADC
Minimum Span: Less than +20 mADC and more than 1 mADC -- outputs can be reversed for both voltage and current outputs

   

Ordering Information
TZ-1CA Configure Input/Output in Store
$337.00
 

Topac Inc. 231 CJC Highway, Suite 103 Cohasset MA 02025 USA
Tel: 781 740 8778 Fax: 781 740 8779
e-mail: sales@topac.com

| Home | E-mail | Online store | Inquiry form | Related |

This is a world wide web document page design layout and HTML code copyright ©2002-2003 by Topac Inc. - ALL RIGHTS RESERVED